Test card
Y-Malachite green speed test reagent
Product
introduction
Application
A. It is widely used to
detect various types of aquarium, ponds
and other water, whether it contains
malachite green ingredients. B. Speed
test
agents malachite green box (Ⅰ)
is only used for preliminary screening
of suspected samples using the malachite
green; C, malachite green speed test
agent box (Ⅱ)
to confirm whether the samples contain
malachite green component.
Test principle
Enriching column can
concentrate and enrich hundred times of
the Malachite green content in the breed
aquatics water, after the malachite
green enriched by the enriching column,
the adsorption in the column will be
some obvious green ribbon. Then the
malachite eluted by the eluant again. If
the eluant appears green that can be
used malachite green. If compare the
eluant with the malachite standard
sample to be characterized, to confirm
whether the use of malachite green.
Specification: 10 times.
Storage: normal
temperature, avoid from sunlight.
Packing: 1. malachite
green enrichment column: 10 branches, 2.
eluent: 2 bottles, 3. the control
solution of malachite green: 1 bottle,
4. ear cleaning balls: 1 (provide for
oneself), 5. Transfusion system: 4 pay.
Period validity: 6
months.
Usage
1. Sampling: First close
the transfusion device at the lower end
of the transfusion pipe flow controller,
using 25ml or 100ml syringe sampling and
want to check the fluid directly from
the infusion bag ( tank ) sample adding
holes to join. When the enriching column
enrichment of 100ml samples can be
detected malachite green minimum
concentration is 3UG / L, 300ml
enrichment can be detected malachite
green the lowest concentration of 1 ug /
L. After the sample is loaded into the
transfusion bag (tank) hanging on a nail
or hook.
2.
Enrichment: (1) Exhaust: According to
the experimental setup, first slowly
open the infusion flow rate controller,
so that the liquid from the drainage
guide Tube out until the bottom of the
liquid outflow from the catheter side by
side under a clean air-water separator
tube in the bubble (the liquid should be
filled with gas-water separator below
the catheter and the catheter should be
absolutely no air bubbles), quick adjust
water flow rate controller is switched
off, rapidly the enrichment column big
mouth on the end of the receiving
drainage catheter. Accordance with the
"speed test case for use of malachite
green," photos two of the "Exhaust" port
up under the enrichment column, and then
open the flow controller, so that slowly
filled with liquid enrichment column,
and the discharge column gas. (2) Such
as "malachite green-speed measuring box
for use" as shown in photo one, put down
the enrichment column, under the port,
so that the liquid Flow through the
enrichment column for enrichment.
Control the flow rate of about 60 drops
per minute.
3. Elution: flow to be
completely liquid sample enrichment
column, the enrichment column off; wash
the ear with the ball towards the column
from the column packing pressure port
blowing, the water column to catch a
clean and dry as possible. Enrichment
column from the column after drying by
adding 1 ml eluant port and let the
liquid naturally shed, collecting eluate
in the control vials.
4. Results of
observation: If cylinder filler layer
that was green and down as the eluent,
the preliminary view that may contain
malachite green, otherwise it is
negative. Or collection of the eluate
can be initially considered for the
green malachite green may be.
Main technical
parameters:
Flow through the
enrichment column sample enrichment time
to detect the lower limit of detection
100ml
25-30min
3.0 ug/L(3.0ppb)
300ml
75 - 90min 1.0 ug/L(1.0ppb)
Announcements
1. If the turbidity test
samples should be filtered before use.
2. On-site detection of
positive samples should be used to test
reagents malachite green-speed box (Ⅱ)
to confirm or high performance liquid
chromatography.
3. The use of malachite
green standard working solution: This
kit provides the standard working
solution available for beginners
training. Standard working solution
concentration is 75ug/ml, before using
to draw 50mL water to adding a drop of
standard working solution that has been
diluted to 0.075ug/mL of malachite green
solution. Take 15ml diluents through the
enrichment column, column packing
enriched significantly in the green.
|